Guangzhou YIGU Medical Equipment Service Co.,Ltd
---Engineer have 12 years maintenance experience
Phone / whatsApp/wechat: +86 13760745557 / +86 13760746611
office:+86 -20-29894177 E-mail:sales@gzyg-med.com
Guangzhou YIGU Medical Equipment Service Co.,Ltd
---Engineer have 12 years maintenance experience
Phone / whatsApp/wechat: +86 13760745557 / +86 13760746611
office:+86 -20-29894177 E-mail:sales@gzyg-med.com
![]() |
Here is some snaps of our stocks,if you want to know more about our production,can text me us freely. We are mainly dealing and repairing patient monitors and parts,our targets customers is all over the word,if you want to be one of them ? We always stay positive and opening to the market,if you ... Read More
|
![]() |
Hi Sir; Thanks for your notice,we are having National Holiday from 1st till 8th Oct in 2020! If any demands pls feel free to contact with us,we will try our best to provide services once we have time during this period. Hope you guys happy all the day! Thanks & Regards Read More
|
![]() |
Dear Yigu Medical Customers; Wish you Happy 2020! We will back to work on 2th January,any demands pls leave us message,we will reply asap! Best Wishes Read More
|
![]() |
2019 Dusseldorf Medical Exhibition This year (2019) it's the first year for yigu medical to attend the fair,we meet some of our clients on the fair. We intorduce each other and get more familiar with each after that. Hope can work with more of you in the near future. Yigu Medical is always warmly ... Read More
|
![]() |
We generally received the machine and found that it could not be turned on. (SIMILAR Repairing Experience) First check the insurance and found that the fuse is good. Check the rectifier bridge later and find that there is also 324V voltage. The disconnection voltage is measured again and the voltage ... Read More
|
![]() |
GE DASH2500 ECG monitor Dark screen Fault description: GE DASH2500 monitor is connected to the power supply, the power indicator light is on, press the power button, you can hear the ‘drop’ boot sound. With the black screen, there is no response by NIBP. Inspection steps: suspected high voltage ... Read More
|
![]() |
Philips HeartStart XL M4735A defibrillator can not be turned on failure repair Symptom: The M4735A defibrillator is connected to the AC power supply. The power indicator is on and cannot be turned on. Inspection steps and maintenance: First suspected that the encoder is faulty, the encoder is ... Read More
|
![]() |
Hi Sir; Thanks for your notice,we are having National Holiday from 1st till 7th Oct. If any demands pls feel free to contact with us,we will try our best to provide services once we have time during this period. Thanks & Regards Read More
|
![]() |
Mid-Autumn Festival, also known as the Moon Festival, the Autumn Festival, the Festival of Sacrificing the Moon, the Mid-Autumn Festival, the Festival of Saluting the Moon, the Festival of Reunion and so on, is a traditional Chinese folk festival. The Mid-Autumn Festival originated from the worship ... Read More
|
![]() |
Malfunction: when the GE Solar8000 monitor measures the blood pressure, (the blood pressure reaches about 90) the blood pressure cannot go up Fault analysis: blood pressure fault is a common fault of the monitor. The main reasons are as follows: (1) leakage of cuff; (2) connection pipeline leakage; ... Read More
|